Reports until 12:51, Monday 21 October 2019
H1 PSL (PSL)
corey.gray@LIGO.ORG - posted 12:51, Monday 21 October 2019 (52595)
PSL Weekly Report (FAMIS #11035)

PSL Report FAMIS report:
    Laser Status:
    Front End Power is 32.5W (should be around 30 W)
    70W Output Power is 69.29W
    Front End Watch is GREEN
    70W Watch is GREEN

    PMC:
    It has been locked 3 days, 20 hr 52 minutes (should be days/weeks)
    Reflected power = 11.37Watts
    Transmitted power = 53.08Watts
    PowerSum = 64.45Watts.

    FSS:
    It has been locked for 2 days 18 hr and 3 min (should be days/weeks)
    TPD[V] = 5.138V (min 0.9V)

    ISS:
    The diffracted power is around 1.9%
    Last saturation event was 3 days 3 hours and 35 minutes ago (should be days/weeks)


    Possible Issues:  None