Displaying report
1
-
1
of
1
.
Reports until 18:44, Tuesday 22 October 2019
LHO VE
Link
kyle.ryan@LIGO.ORG -
posted 18:44, Tuesday 22 October 2019
(52634)
YBM turbo installation -> Found 2 x 10-8 torr*L/sec leak (helium) at the rotatable turbo fixation/6-way cross joint
Gerardo M., Kyle R.
Bummer! We were so close! We'll start the day tomorrow addressing this leak.
Displaying report
1
-
1
of
1
.