LSC AA Chassis Modified

WP 8375
E1900130
E1900105

Three of the four LSC AA chassis in the corner station have been modified per ECR1900105/E1900130. Input buffer U2 was replaced on all channels with faster slew rate opamp ADA4622.

S1102784 S1301308 S1102789

F. Clara, E. Merilh, M Pirello