Reports until 09:01, Saturday 26 October 2019
H1 CDS (CDS, VE)
sheila.dwyer@LIGO.ORG - posted 09:01, Saturday 26 October 2019 - last comment - 10:46, Saturday 26 October 2019(52698)
HAM6 high votlage off

The HAM6 shutter doesn't have high voltage.  (The medm screen says that the high votlage is OK, but the LED on the driver says the capactior is only charged to 13V.) With Varun as a phone buddy I went to the mezzanine, the HAM6 high voltage supply is off.  Attached is a trend of PT-110 over night, there is a spike to 4e-5 Torr around midnight .  It is reading 1.1e-6 Torr now.

Images attached to this report
Comments related to this report
sheila.dwyer@LIGO.ORG - 09:26, Saturday 26 October 2019 (52699)

Talked to Richard on the phone and reset the high voltage.

kyle.ryan@LIGO.ORG - 10:18, Saturday 26 October 2019 (52701)

On second reading, this is another example of the HV tripping off due to the gauge noise spike observed earlier this week.  As discussed then, we can swap-out the gauge electronics as the first step in a troubleshooting process.  

kyle.ryan@LIGO.ORG - 10:46, Saturday 26 October 2019 (52702)

Gerardo had suggested degassing the hot filament last week but I don't see a LOG entry recording this.  This should be done before anything else.  I think that Patrick T. can do this via CDS?  Beckhoff?