Displaying report 1-1 of 1.
Reports until 16:24, Tuesday 29 October 2019
LHO VE
chandra.romel@LIGO.ORG - posted 16:24, Tuesday 29 October 2019 - last comment - 17:50, Tuesday 29 October 2019(52781)
RGA scans of corner: pre & post vent

RGA scans were collected before and after venting the corner station in October. All scans are in SEM mode and attached with notes below.

I was concerned about the HC signature shown in Oct. 23 post-vent scan, so today I cycled GV 1,2,5,7 to isolate a potential problem; all scans look satisfactory compared to Oct. 23. The BSC 7,8 doors were cleaned on site, air baked and FTIR sampled. Today we received the pre-bake, post-clean FTIR results from JPL (E1900244) showing some low level residue, but mostly within the LIGO spec of 0.02 microgram/cm^2. The non-vacuum threaded holes used for rigging are extremely contaminated even after Kyle and Chris' best attempt at cleaning them with wand, due to vendor accidentally using silver anti-seize grease. Note that the serial numbers of plates were accidentally stamped on vacuum side. SN001 is on BSC8 and SN002 is on BSC7.

Both plates were sampled with FTIR kit after air baking at 150C for 48 hours; results from JPL were within LIGO specs (<0.02 microgram/cm^2) and received before installing plates (E1900248).

Based on pre and post vent RGA scans from October vent, I conclude that the two scans that look bad are because the filament was not fully warmed up. It should be energized for at least an hour before scanning.

 

Non-image files attached to this report
Comments related to this report
chandra.romel@LIGO.ORG - 17:50, Tuesday 29 October 2019 (52785)

GV 1,2,5,7 were soft closed at various times starting at 9am during today's experiment and RGA scans collected about 1/2 hr after each valve change (filament has been on since yesterday at least). RGA is located between HAM 4 and 5. GV 1,2 took 4:39 min. to soft close (48" valves). (44" valves take 4:18-4:23 to soft close). GV 5,7 soft closed with 5 psig pressure applied to piston.