Kyle prepared a new Prisma Plus RGA for HAM6 chamber by baking it remotely prior to installation during Oct. vent. Background is pristine and includes a 10 l/s ion pump to preserve vacuum during chamber vents and also two calibration gas bottles.
Yesterday we had a chance to finally scan HAM6 chamber at total pressure of 9e-7 Torr. RGA was in SEM mode and filament was warmed up for several hours. Scan attached.
Note that the RGA assembly's 10 l/s ion pump is on (using as interium pressure gauge while PT110 is offline) here and has a comparable net pump speed as the conductance to HAM6. As such, a factor of ~2 bias exists. If it were off, as would be nominally the case, the peaks would be ~twice as large as shown.