Reports until 11:54, Wednesday 30 October 2019
LHO VE
chandra.romel@LIGO.ORG - posted 11:54, Wednesday 30 October 2019 - last comment - 10:12, Thursday 31 October 2019(52810)
HAM6 RGA scan

Kyle prepared a new Prisma Plus RGA for HAM6 chamber by baking it remotely prior to installation during Oct. vent. Background is pristine and includes a 10 l/s ion pump to preserve vacuum during chamber vents and also two calibration gas bottles.

Yesterday we had a chance to finally scan HAM6 chamber at total pressure of 9e-7 Torr. RGA was in SEM mode and filament was warmed up for several hours. Scan attached.

Images attached to this report
Non-image files attached to this report
Comments related to this report
kyle.ryan@LIGO.ORG - 10:12, Thursday 31 October 2019 (52831)

Note that the RGA assembly's 10 l/s ion pump is on (using as interium pressure gauge while PT110 is offline) here and has a comparable net pump speed as the conductance to HAM6.  As such, a factor of ~2 bias exists.  If it were off, as would be nominally the case, the peaks would be ~twice as large as shown.