TITLE: 11/04 Day Shift: 16:00-00:00 UTC (08:00-16:00 PST), all times posted in UTC
STATE of H1: Observing at 115Mpc
INCOMING OPERATOR: Patrick
SHIFT SUMMARY: H1 in Observe, instrument air pumps for LVEA are working, SQZ just fixed itself, causing an increase in range
LOG:
VACUUM:
- 15:06:24 - first vacuum alarm for PT199
- 15:19:42UTC - JeffB turns on the pump(s)
- 15:20:16UTC - JeffB turns off the pump(s)
- 15:25:44UTC - preassure levels off at 41.7PSIG
- 15:28:31UTC - JeffB turns the pump(s) on, at 41.6PSIG
- 15:43:15UTC - JeffB turns the pump(s) off
- 15:54:09UTC - Kyle is here, turns the pumps on, pressure 75.2PSIG
- 16:02:290UTC - Kyle turns the pumps off, valves out the LVEA, pressure 98PSIG
- 17:01:12UTC - Kyle turns the pumps back on, valves the LVEA back in, pressure 89PSIG
- 17:23:57UTC - pressure is at lowest point, 62.9PSIG, and pumps come back on
- since then pumps running between 62PSIG and 88PSIG
SQZ:
- 17:24UTC - H1 range drops
- 17:46UTC - H1 out of Observe, I take all SQZ guardians (except beatnote) through INIT to DOWN
- 17:48UTC - SQZ is relocked, no change in range
- 17:56UTC - SQZ guardians to down, cleared SQZ ASC, would not recommend
- 18:13UTC - SQZ relocks, after I restore ZM1 and ZM2 pointing, not exactly, but close
- 18:17UTC - H1 back in Observe
- 23:44UTC - H1 range goes up, ZM1 yaw changes direction
Images that go with the PT199 timeline:
An interesting plot showing an alignment change in ZM2 yaw that coincides with the increase in H1 Range.
The LASSO results for 4 Nov (see https://ldas-jobs.ligo-wa.caltech.edu/~detchar/summary/day/20191104/detchar/lasso/) show a strong correlation between SQZ channels (perhaps excited by seismic motion) and the range drop. Similar results exist for 1 Nov, the most recent prior day with a long lock crossing the range drop.