vincent.lhuillier@LIGO.ORG - posted 18:03, Monday 28 January 2013 (5295)
TF - BS - T240s (corners 2-3) not working right

Transfer functions were measured during the week-end. Stage 1 and stage 2 look good (cf TF_L4C & TF_GS13). On the other hand, the TFs from stage 1 actuators to T240s in corner 2 and corner 3 don't look right (gain and resonance). These 2 instruments don't seem to work properly.

Images attached to this report