rahul.kumar@LIGO.ORG - posted 10:58, Tuesday 05 November 2019 - last comment - 11:37, Tuesday 05 November 2019(53007)
OPLEV charge measurements on ETMX and ETMY

Attached below are the results for the charge measurements performed on the ETMX and ETMY (post Oct 2019 vent break).

ETMX: There has been a definite rise in the charge accumulations on all 4 quadrant for the pitch and yaw mode. The first and the 4th quadrant for the pitch mode shows an average bias voltage of +40V and +20V respectively. The  bias voltage on the 2nd, 3rd and 4th quadrant for the pitch mode has jumped by 10-20V.  For the Yaw mode, the 1st and the 3rd quadrant are at 50V, which is quite high. The 3rd quadrant is just above 40V.

I will discuss these results with J. Kissel to see if we need to flip the sign yet.

 

ETMY: There hasn't been a huge change and the effective bias voltage on most of the quadrant has remained the same or declined. The biggest fall has been seen on the 2nd quadrant (Yaw). The rest of them are between 0-20V.

The slider values were restored after the measurement was complete, and any SDF difference was removed.

 

Images attached to this report
Comments related to this report
rahul.kumar@LIGO.ORG - 11:37, Tuesday 05 November 2019 (53011)

adding a plot of time varying calibrations lines and the Kappa values are off by say 2%. Just for comparison from Sept, the then Kappa values were at 1.0

Images attached to this comment