Found the Y-end MTP shut down during the night -> Problem turned out to be the "over head" chilled water supply valve to the mechanical room was closed for some reason -> Opened valve and restarted MTP -OK Upon measuring the vent/purge air dew point just prior to venting I discovered a problem -> Found that the left drying tower wasn't regenerating correctly -> problem ended up being a wore out swing-style check valve -> Replaced internal (wear) components -> OK -> These check valves (4 per vent/purge skid) have been trouble-free for the first fifteen years of operation but are now failing regularly due to excessive wear -> I will attempt to replace these site wide time permitting. ~1530 hrs. local Began "slow" vent of Y-end ~1600 hrs. local Made small vent rate adjustment ~1630 hrs. local Made a second small vent rate adjustment