Displaying report 1-1 of 1.
Reports until 16:45, Friday 08 November 2019
LHO VE
chandra.romel@LIGO.ORG - posted 16:45, Friday 08 November 2019 (53108)
new PT-110 pressure gauge still noisy

The pressure gauge on HAM6 was replaced due to noisy read back (spiked above high voltage pressure set point more than once causing fast shutter to close) and then gauge failure during troubleshooting (pirani or filament gone bad? - gauge sent back to manufacturer via RMA).

We replaced the entire gauge but the readback is still noisy. Suggest we leave the HV interlock on the new ion pump controller at HAM6 until we can trend gauge over longer period of time to confirm it won't spike above 1e-5 Torr. Not sure what is causing the spikes.

Images attached to this report