Summary: Back to Observing with 1st lock, and only needed intervention for: x & y green arms & Finding IR.
- 9:03 lockloss!
- LOCK#1 9:05 - 10:12
- Green arms: Y-arm not locking. X-arm locked with TEM00, but barely above 0.6.
- After 5-min Ops Wait: ETMx Pit moved from 65.1 to 65.7 (then WFS optimize). ETMy Pit moved from 120.5 to 122.3 (until WFS optimize).
- 9:15-9:18 IR Not Found. Usual Strip Tool was replaced by ndscope on front wall today, and I could not read situation on ndscope because signals aren't scaled to what we are used to(!). So I opened up the Strip Tool on nuc30. Manual Intervention: sweeping ALS Diff Offset to optimize IR Y arm power.
- 10:12 Nominal Low Noise/Observing