Displaying report 1-1 of 1.
Reports until 13:46, Tuesday 26 November 2019
H1 SQZ
daniel.sigg@LIGO.ORG - posted 13:46, Tuesday 26 November 2019 (53500)
Reduced the CLF power

Reduced the CLF power incident on the CLF PD to a nominal 5µW. The CLF waveplate was adjusted, so the that CLF ISS drive offset stayed near 2.7V. This required the following changes:

channel new old
H1:SQZ-CLF_ISS_SETPOINT 0.130 0.510
H1:SQZ-CLF_ISS_GAIN 26 7
H1:SQZ-CLF_SERVO_IN2GAIN +16 –4
H1:SQZ-OPO_IR_RESONANCE_CLF_NOM 0.005 0.05
H1:SQZ-LO_SERVO_IN1GAIN +8 –5

The CLF ugf was measured to be 2.2kHz, the CLF ISS ugf is 10kHz, and the LO ugf is 4kHz.

Displaying report 1-1 of 1.