Looks like a general core crash, models remain running. Unlike earlier h1seib3 crash, in this case no other systems are affected and only DAQ data from SEI-B1 is down.
I've taken h1seib1 out of the Dolphin fabric, which caused lock loss. Ed is now rebooting the machine via the front panel reset button.
Opened FRS13892
Computer rebooted with no problems. No IRIG-B timing issue. All ADC and DAC cards seen.
Over to Ed for IFO recovery