TEST OFFSETs on all optics are typically a fixed value, and not active during NLN/Observe. Currentl there are two situations that need correction/understanding, listed below:
- This weekend with reboots, ETMX TEST Yaw Offset became 0, when nominal is -8.4. This was acknowledged by Patrick, and is on our list to revert. This does not change the alignment of ETMX, no functional change to H1 in NLN, may have caused issues with Initial Alignment.
- TMS TEST Offset values have been changing since 19 Nov 2019. This is different than before that date, where the values were fixed. The offsets are currently ON, which means they are involved with the overall alignment of TMS. This is different than all ofther optics. Currently working to understand the change.
Plots of ETMX TEST YAW and TMS TEST OFFSETs attached.
Current list of TEST OFFSET values:
- H1:SUS-BS_M1_TEST_P_OFFSET = 12.6
- H1:SUS-BS_M1_TEST_Y_OFFSET = -19.2
- H1:SUS-ETMX_M0_TEST_P_OFFSET = 10.1
- H1:SUS-ETMX_M0_TEST_Y_OFFSET = -8.4, TO BE REVERTED, should be -8.4, currently 0
- H1:SUS-ETMY_M0_TEST_P_OFFSET = 10.1
- H1:SUS-ETMY_M0_TEST_Y_OFFSET = 8.4
- H1:SUS-IM1_M1_TEST_P_OFFSET = 333.0
- H1:SUS-IM1_M1_TEST_Y_OFFSET = 333.0
- H1:SUS-IM2_M1_TEST_P_OFFSET = -700.0
- H1:SUS-IM2_M1_TEST_Y_OFFSET = -520.0
- H1:SUS-IM3_M1_TEST_P_OFFSET = 555.0
- H1:SUS-IM3_M1_TEST_Y_OFFSET = 555.0
- H1:SUS-IM4_M1_TEST_P_OFFSET = 555.0
- H1:SUS-IM4_M1_TEST_Y_OFFSET = 555.0
- H1:SUS-ITMX_M0_TEST_P_OFFSET = -300.0
- H1:SUS-ITMX_M0_TEST_Y_OFFSET = -16.9
- H1:SUS-ITMY_M0_TEST_P_OFFSET = 400.0
- H1:SUS-ITMY_M0_TEST_Y_OFFSET = 100.0
- H1:SUS-MC1_M1_TEST_P_OFFSET = 0.0
- H1:SUS-MC1_M1_TEST_Y_OFFSET = 0.0
- H1:SUS-MC2_M1_TEST_P_OFFSET = 0.0
- H1:SUS-MC2_M1_TEST_Y_OFFSET = -1000.0
- H1:SUS-MC3_M1_TEST_P_OFFSET = 2000.0
- H1:SUS-MC3_M1_TEST_Y_OFFSET = 2000.0
- H1:SUS-OFI_M1_TEST_T_OFFSET = 0.0
- H1:SUS-OFI_M1_TEST_Y_OFFSET = 0.0
- H1:SUS-OM1_M1_TEST_P_OFFSET = 0.0
- H1:SUS-OM1_M1_TEST_Y_OFFSET = 0.0
- H1:SUS-OM2_M1_TEST_P_OFFSET = 0.0
- H1:SUS-OM2_M1_TEST_Y_OFFSET = 0.0
- H1:SUS-OM3_M1_TEST_P_OFFSET = 800.0
- H1:SUS-OM3_M1_TEST_Y_OFFSET = 1000.0
- H1:SUS-OMC_M1_TEST_P_OFFSET = 100.0
- H1:SUS-OMC_M1_TEST_Y_OFFSET = 20.0
- H1:SUS-OPO_M1_TEST_P_OFFSET = 0.0
- H1:SUS-OPO_M1_TEST_Y_OFFSET = 0.0
- H1:SUS-PR2_M1_TEST_P_OFFSET = 500.0
- H1:SUS-PR2_M1_TEST_Y_OFFSET = 500.0
- H1:SUS-PR3_M1_TEST_P_OFFSET = -705.7
- H1:SUS-PR3_M1_TEST_Y_OFFSET = -3338.0
- H1:SUS-PRM_M1_TEST_P_OFFSET = 681.0
- H1:SUS-PRM_M1_TEST_Y_OFFSET = -5914.0
- H1:SUS-RM1_M1_TEST_P_OFFSET = 0.0
- H1:SUS-RM1_M1_TEST_Y_OFFSET = 0.0
- H1:SUS-RM2_M1_TEST_P_OFFSET = 0.0
- H1:SUS-RM2_M1_TEST_Y_OFFSET = 100.0
- H1:SUS-SR2_M1_TEST_P_OFFSET = 200.0
- H1:SUS-SR2_M1_TEST_Y_OFFSET = 300.0
- H1:SUS-SR3_M1_TEST_P_OFFSET = -555.0
- H1:SUS-SR3_M1_TEST_Y_OFFSET = -152.3
- H1:SUS-SRM_M1_TEST_P_OFFSET = -1400.0
- H1:SUS-SRM_M1_TEST_Y_OFFSET = 1290.0
- H1:SUS-TMSX_M1_TEST_P_OFFSET = -8.6, ALL TMS values were fixed until 19 Nov 2019
- H1:SUS-TMSX_M1_TEST_Y_OFFSET = 31.6
- H1:SUS-TMSY_M1_TEST_P_OFFSET = 40.1
- H1:SUS-TMSY_M1_TEST_Y_OFFSET = 34.6
- H1:SUS-ZM1_M1_TEST_P_OFFSET = 0.0
- H1:SUS-ZM1_M1_TEST_Y_OFFSET = 0.0
- H1:SUS-ZM2_M1_TEST_P_OFFSET = 0.0
- H1:SUS-ZM2_M1_TEST_Y_OFFSET = 0.0