Displaying report 1-1 of 1.
Reports until 17:47, Monday 02 December 2019
H1 PSL
cheryl.vorvick@LIGO.ORG - posted 17:47, Monday 02 December 2019 (53636)
PSL Weekly Report, FAMIS 11041

    Laser Status:
    Front End Power is 32.33W (should be around 30 W)
    70W Output Power is 69.29W
    Front End Watch is GREEN
    70W Watch is GREEN

    PMC:
    It has been locked 27 days, 6 hr 55 minutes (should be days/weeks)
    Reflected power = 11.59Watts
    Transmitted power = 52.38Watts
    PowerSum = 63.97Watts.

    FSS:
    It has been locked for 0 days 10 hr and 55 min (should be days/weeks)
    TPD[V] = 2.746V (min 0.9V)

    ISS:
    The diffracted power is around 2.5%
    Last saturation event was 0 days 10 hours and 55 minutes ago (should be days/weeks)


    Possible Issues:

    TPD[V] = 2.746V (min 0.9V) - Jason will be addressing this during Maintenance tomorrow

Displaying report 1-1 of 1.