Reports until 05:24, Tuesday 03 December 2019
H1 SEI
jeffrey.bartlett@LIGO.ORG - posted 05:24, Tuesday 03 December 2019 (53642)
SEI Seismometer Mass Position Check (FAMIS #9248)
   Ran the STS Centering and T240 Centering check scripts. 

   STS: 2019-12-03 05:18:06.357834

There are 1 STS proof masses out of range ( > 2.0 [V] )!
STS B DOF X/U = 4.702 [V]

All other proof masses are within range ( < 2.0 [V] ):
STS A DOF X/U = -0.757 [V]
STS A DOF Y/V = -1.084 [V]
STS A DOF Z/W = -0.264 [V]
STS B DOF Y/V = -0.495 [V]
STS B DOF Z/W = -0.345 [V]
STS C DOF X/U = 0.344 [V]
STS C DOF Y/V = 0.882 [V]
STS C DOF Z/W = -0.11 [V]
STS EX DOF X/U = -0.176 [V]
STS EX DOF Y/V = 0.231 [V]
STS EX DOF Z/W = 0.265 [V]
STS EY DOF X/U = 0.397 [V]
STS EY DOF Y/V = -0.326 [V]
STS EY DOF Z/W = 0.693 [V]


Assessment complete.

T240: 2019-12-03 05:18:30.363749


There are 8 T240 proof masses out of range ( > 0.3 [V] )!
ETMX T240 2 DOF X/U = -0.939 [V]
ETMX T240 2 DOF Y/V = -1.005 [V]
ETMX T240 2 DOF Z/W = -0.575 [V]
ITMX T240 1 DOF X/U = -1.212 [V]
ITMX T240 3 DOF X/U = -1.227 [V]
ITMY T240 2 DOF Y/V = 0.3 [V]
ITMY T240 3 DOF X/U = -0.43 [V]
ITMY T240 3 DOF Z/W = -1.541 [V]


All other proof masses are within range ( < 0.3 [V] ):
ETMX T240 1 DOF X/U = 0.03 [V]
ETMX T240 1 DOF Y/V = -0.038 [V]
ETMX T240 1 DOF Z/W = -0.111 [V]
ETMX T240 3 DOF X/U = -0.101 [V]
ETMX T240 3 DOF Y/V = -0.058 [V]
ETMX T240 3 DOF Z/W = -0.05 [V]
ETMY T240 1 DOF X/U = -0.116 [V]
ETMY T240 1 DOF Y/V = 0.251 [V]
ETMY T240 1 DOF Z/W = -0.048 [V]
ETMY T240 2 DOF X/U = -0.029 [V]
ETMY T240 2 DOF Y/V = -0.086 [V]
ETMY T240 2 DOF Z/W = -0.08 [V]
ETMY T240 3 DOF X/U = -0.061 [V]
ETMY T240 3 DOF Y/V = -0.118 [V]
ETMY T240 3 DOF Z/W = 0.115 [V]
ITMX T240 1 DOF Y/V = 0.146 [V]
ITMX T240 1 DOF Z/W = 0.153 [V]
ITMX T240 2 DOF X/U = 0.132 [V]
ITMX T240 2 DOF Y/V = 0.214 [V]
ITMX T240 2 DOF Z/W = 0.219 [V]
ITMX T240 3 DOF Y/V = 0.111 [V]
ITMX T240 3 DOF Z/W = -0.012 [V]
ITMY T240 1 DOF X/U = 0.17 [V]
ITMY T240 1 DOF Y/V = 0.145 [V]
ITMY T240 1 DOF Z/W = 0.22 [V]
ITMY T240 2 DOF X/U = 0.107 [V]
ITMY T240 2 DOF Z/W = 0.236 [V]
ITMY T240 3 DOF Y/V = 0.209 [V]
BS T240 1 DOF X/U = 0.014 [V]
BS T240 1 DOF Y/V = -0.093 [V]
BS T240 1 DOF Z/W = 0.298 [V]
BS T240 2 DOF X/U = 0.095 [V]
BS T240 2 DOF Y/V = 0.271 [V]
BS T240 2 DOF Z/W = 0.013 [V]
BS T240 3 DOF X/U = 0.09 [V]
BS T240 3 DOF Y/V = -0.149 [V]
BS T240 3 DOF Z/W = -0.162 [V]


Assessment complete.


   Closing FAMIS #9248