TITLE: 12/12 Eve Shift: 00:00-08:00 UTC (16:00-00:00 PST), all times posted in UTC
STATE of H1: Observing at 110Mpc
INCOMING OPERATOR: Corey
SHIFT SUMMARY: locked in Observe most of the shift
LOG:
- 00:18:22 - H1 in Observe
- 00:35:00 - LVEA lights were turned off
- at 00:35:00, Gerardo opened the High Bay door the the LVEA, and turned off the LVEA lights
- the lights in the Clean Area and CER were also turned off (before the LVEA lights)
- Gerardo did not enter the LVEA, so did not walk on the Technical Slab
- 00:18-00:35 Observe mode was non-standard, due to lights and access to the above areas
- 01:14 - Kyle to MY
- 01:31 - Kyle back from MY
- nuc25 had to be restarted
- ITMX violin mode damping has non-standard gains
- IH1:SUS-ITMX_L2_DAMP_MODE17_GAIN, 1003.12Hz, nominal = 20, gain = 120, higher gain needed to damp the peak
- H1:SUS-ITMX_L2_DAMP_MODE19_GAIN, 998.018Hz, nominal = 0, gain = 60, usually I use gain of 20, higher gain needed to damp the peak
- ETMY violin mode damping has non-standard gains
- H1:SUS-ETMY_L2_DAMP_MODE11_GAIN, 1000.061Hz, nominal = 150, gain = 0, peak was damped, no longer visible in spectra
- H1:SUS-ETMY_L2_DAMP_MODE12_GAIN, 1000.423Hz, nominal = 2, gain = 0, peak is visible, but small, and damping interfeared with modes 18 and 20
- H1:SUS-ETMY_L2_DAMP_MODE13_GAIN, 1009.932Hz, nominal = 130, gain = 60, peak was no longer visible in spectra, then it was again
- used a reduced gain to avoid interfering with modes 18 and 20
- H1:SUS-ETMY_L2_DAMP_MODE14_GAIN, 1017.985Hz, nominal = 100, gain = 50, peak was no longer visible in spectra, then it was again
- used a reduced gain to avoid interfering with modes 18 and 20
- H1:SUS-ETMY_L2_DAMP_MODE15_GAIN, 1018.238Hz, nominal = 1500, gain = 0, peak was damped, no longer visible in spectra
- H1:SUS-ETMY_L2_DAMP_MODE18_GAIN, 1000.296Hz, nominal = 0, gain = -8,
- with modes 12 and 20 no longer visible in spectra, and damping turned off, allowed me to increase mode 18 gain to -8
- H1:SUS-ETMY_L2_DAMP_MODE20_GAIN, 1000.307Hz, nominal = 1, gain = 0, peak was damped, no longer visible in spectra
- I've asked Corey to watch ITMX and ETMY violins, since these damping gains are set up to damp ETMY mode 18, and not standard config.