H1 lost lock, and relocked with the old ETMY violin mode damping gains, which, after 10 hours, were ringing up modes 18 and 20.
I currently have the ETMY violin mode gains to these values:
- mode 11: was 150, now 50
- mode 12, was 2, now 0.5
- mode 18, was 0, now -10
- mode 20, was 1.0, now 0.5
- these gains have been stable for around an hour
I changed lscparams to use the above gains if H1 loses lock
- in lscparams
- mode 11: was 150, now 50
- mode 12, was 2, now 0.5
- mode 18, was 0, now 10, plus I engaged the filter FM6, which flips the sign
- this combination matches the -10 gain I'm using in the current lock
- mode 20, was 1.0, now 0.5
- violin guardian reload needed, if H1 loses lock
Plots attached show the damping done tonight.