H1 CAL
jeffrey.kissel@LIGO.ORG - posted 17:30, Wednesday 18 December 2019 (53980)
Calibration Measurements: Most of a Full Suite Today Before EQ, Updates to Sensing Function
J. Kissel

I've captured a standard sensing function suite, and was able to get in a full L3 and some on an L2 actuator sweep today. Attached are updates to the sensing function collection of measurements. These thermalized data sets are showing *pretty* consistent results. I'm working on getting more channels from the frames (PSL input power, temperature in the LVEA) to see if I can find other things that can predict the trends I see in f_s^2. One idea I had was that the steps between lock stretches are correlated to whether an initial alignment was run. I think I can conclude that there's no correlatino there. We're also working on backing out how bad things are *during* thermalization, with the data gathered yesterday (LHO aLOG 53951). 

Raw data from today:
Sensing Function:
/ligo/svncommon/CalSVN/aligocalibration/trunk/Runs/O3/H1/Measurements/FullIFOSensingTFs/
    2019-12-18_H1_DARM_OLGTF_LF_SS_5to1100Hz_15min.xml
    2019-12-18_H1_PCALY2DARMTF_LF_SS_5t1100Hz_10min.xml
    2019-12-18_H1_PCALY2DARMTF_BB_3min.xml

Actuation Function:
/ligo/svncommon/CalSVN/aligocalibration/trunk/Runs/O3/H1/Measurements/FullIFOActuationTFs/
    2019-12-18_H1SUSETMX_L3_iEXC2DARM_12min.xml
    2019-12-18_H1SUSETMX_L3_PCAL2DARM_6min.xml
    2019-12-18_H1SUSETMX_L2_iEXC2DARM_12min.xml
    2019-12-18_H1SUSETMX_L2_PCAL2DARM_6min.xml    << only half a measurement, because of the canadian EQ

As always -- more to come!