rahul.kumar@LIGO.ORG - posted 11:45, Monday 23 December 2019 - last comment - 12:55, Monday 23 December 2019(54066)
ETMX mode 6

Dave, Corey, Rahul

Yesterday, Cheryl made changes on the ETMX mode 6 damping filter settings (alog 54043), which she did not load to make sure that we don't go out of observing. This morning I loaded these changes (alog 54063) . However, it didnt go well and etmx mode 6 got excited very quickly (monitor level 5.5ish), after IFO was relocked (lockloss due to an earthquake) this morning and new settings were initiated. I manually adjusted the gain (reduced it from +5 to +1) which worked and the peak excitation went below 1.0. However according to Cheryl's alog the changes would also affect ETMY mode 6 (510.7264Hz) - which I still have to understand how.

In the meanwhile to keep things simple and buy some time, I requested Dave to revert this specific change and he will be posting a separate alog on this.

 

Images attached to this report
Comments related to this report
david.barker@LIGO.ORG - 12:55, Monday 23 December 2019 (54070)

I checked Cheryl's MODE6 modified H1SUSTEMX.txt as svn r20778. I then restored the filter file from last Wednesday which was  loaded into h1susetmx. This file has been checked in as r20779.