Reports until 05:40, Wednesday 25 December 2019
H1 PSL (PSL)
yannick.lecoeuche@LIGO.ORG - posted 05:40, Wednesday 25 December 2019 (54107)
PSL Status Report-Weekly

    Laser Status:
    Front End Power is 32.31W (should be around 30 W)
    70W Output Power is 69.36W
    Front End Watch is GREEN
    70W Watch is GREEN

    PMC:
    It has been locked 7 days, 17 hr 24 minutes (should be days/weeks)
    Reflected power = 11.59Watts
    Transmitted power = 51.95Watts
    PowerSum = 63.55Watts.

    FSS:
    It has been locked for 0 days 5 hr and 52 min (should be days/weeks)
    TPD[V] = 5.517V (min 0.9V)

    ISS:
    The diffracted power is around 2.5%
    Last saturation event was 0 days 5 hours and 52 minutes ago (should be days/weeks)


    Possible Issues: