Reports until 13:32, Friday 03 January 2020
H1 AOS
vladimir.bossilkov@LIGO.ORG - posted 13:32, Friday 03 January 2020 - last comment - 13:45, Friday 03 January 2020(54264)
Update to H1CALCS filters - Quantifying change to response function R, and how much improvement to expect in calibration

The soon to be made change to LHO calibration is planned for next Wednesday (20200108).

As a follow up to what the newly modeled UIM dynamics will change I have generated a couple more plots. Previously I looked at how the pyDARM calibration will change with the updated TF (the comment in a related report), highlighting the improvements to calibration, and how they line up with known discrepancies between DELTAL and Pcal.

Now I look at the difference between the new front end, foton filterbank, and the pyDARM calibration contributions to R (2020-01-03_H1_Rfoton_new_over_Rpydarm_new.pdf) - here you see that I am missing two features, one at ~90 Hz and at 134 Hz (you can see these in the comment to the related report as well).

Finally I look at the total change in the front end, as a change in contribution to R (2020-01-03_H1_Rfoton_new_over_Rfoton_old.pdf) . This is equivalent to the change in calibration on the DELTAL.

 

There is quite a lot of alterations at lower frequencies (<40) in the comparison of contributions from the foton filters. Other than the testmass correction and new 50-150Hz UIM features, nothing is supposed to impact these low frequencies. I suspect the reason is that in propagating the the SUSmodels (L1, L2 and L3 stage, i.e. the UIM, PUM and TST stages) to the foton filterbanks, and I optimising the minreal functions of matlab to keep absoluteley as many poles and zeros as I can get away with. For example the UIM stage (where I took the most time and care), is comprised of (ignoring the 1/f^6 filter bank and the gain only bank) 3 filter banks of 10, 10, and 9 Second Order Sections (the max allowed in foton is 10). I used this philosophy for the lower stages also - and I suspect the low frequency parts of the SUSmodels are better represented in this iteration of the filter - giving rise to corrective changes below 40Hz in 2020-01-03_H1_Rfoton_new_over_Rfoton_old.pdf.

Non-image files attached to this report
Comments related to this report
vladimir.bossilkov@LIGO.ORG - 13:45, Friday 03 January 2020 (54266)

I have prepeared a new parameter file:

/ligo/svncommon/CalSVN/aligocalibration/trunk/Runs/O3/H1/params/

modelparams_H1_20200103.py

Which reference to updated H1susdata file:

/ligo/svncommon/CalSVN/aligocalibration/trunk/Common/pyDARM/matlab_scripts/

20200103_H1_03_susdata.mat

I will write a bash script next week that automates the generation of that .mat file, since presently it is a two step process, between matlab and python