Displaying report 1-1 of 1.
Reports until 11:00, Wednesday 08 January 2020
H1 SUS
cheryl.vorvick@LIGO.ORG - posted 11:00, Wednesday 08 January 2020 (54367)
ITMY violin mode 18 damped overnight

Plot attached shows that the violin peak at 998.828Hz, ITMY Mode 18, was damped overnight with a gain of +300.  The nearby peak at 998.961Hz, ITMY Mode 17, was not rung up.

Attachments:

  1. spectra: left peak is IY mode18, right peak is IY mode 17,
    • left plot: REF(green) 25 Nov 2019, cursor values (magenta) 31 Dec 2019
    • right plot: REF(green) 25 Nov 2019, cursor values (red) 8 Jan 2020
  2. overnight trends:
    • top plot: IY mode18 gain and drive
    • middle plot: IY mode18 monitor filter (polluted by mode17)
    • bottom plot: IY mode17 monitor filter (polluted by mode18)
  3. trends since Oct 2019:
    • top plot: IY mode18 gain and drive - mostly zero, not damped
    • middle plot: IY mode18 monitor filter (polluted by mode17) - shows the mode peak increasing over the 2 or so months since the October break
    • bottom plot: IY mode17 monitor filter (polluted by mode18)
Images attached to this report