Lock Loss 2250 UTC

Human caused. We dropped out of observing from a channel connection error when some background violin mode channels were changed. Surprised by this, while we were fixing the issue I allowed some other filters to be loaded, which killed the lock.

Comments related to this report
rahul.kumar@LIGO.ORG - 15:12, Wednesday 08 January 2020 (54374)

Cheryl made changes to the ITMX damping filters for a good reason, however we forgot to reduce the gain to zero before applying those filters. This resulted in an unfortunate lockloss due to a sharp (before we could react) increase in the ITMX violin mode amplitude.