H1 SUS
cheryl.vorvick@LIGO.ORG - posted 03:29, Thursday 09 January 2020 (54378)
ITMX violin modes rung up, lockloss, mode2 re-enabled with output +/-20000, 2 seconds after lockloss

Lockloss details:

I don't know if this is by design, but I found that the violin damping continues for almost 2 seconds:

The mode2 switch must get switched in the DOWN script, because the mode2 output was re-enabled, and what doesn't make sense is that even with the mode2 gain at "0," mode2 output damps for about 5.5 seconds, and the output signal isn't small, it's +/-20000 counts!

Many plots that go with the timeline.

The cause was human error, the response of all the modes continuing to damp may be intentional, the response of mode2 may be understood, but was unexpected when I looked at the timeline of the signals.