New BS model
Vincent was keen to test IPC between his new BSC2 ISI model and the BS SUS model, but the latter hadn't been created so I waded in and created it. I adapted the existing BSTST model and then compared it at the end with the L1 version - they were gratifyingly similar, except for the IPC stuff, which only exists in the H1 version so far. I got it running and updated the sitemap to point to it. I installed projection matrices, filters, and default gains and offsets, and created a safe.snap. 

Along the way I discovered that the SUS_CUST_BSFM_ADC_MONITOR.adl screen had been hardwired to L1, so I fixed that.

I noticed a few anomalies that need further debugging:

* The output fields on the SUS_CUST_BSFM_OVERVIEW.adl screen between the USER DACKILL and the IOP DACKILL (H1:FEC-31_DAC_OUTPUT_2_4 etc) are white. There's also whiteness in the H1CDSOVERVIEW.adl screen.

* The M2LR OSEM is showing a non-zero, fluctuating value (≈±2500 counts) despite having no OSEM connected.

* There's red in the Dolphin IPC bit on the H1CDSOVERVIEW.adl screen.

Apart from all of that(!), very possibly when the physical suspension is connected up it will damp.