While RGA manufacturer was on site testing Labview-based RGA software, we noticed a significant AMU 28 peak at EX, towering water signature, suggesting an air leak. The total pressure is 1.9e-9 Torr and the trend is relatively flat, so no urgent concern. I suspect this is due to the failed annulus ion pump on GV20. We may have an inner o-ring leak on GV20 in addition to diffusion through o-rings.
AIP has been off for over a week as EX has been inaccessable due to tumbleweed blockages.
EY scan is also attached for comparison.