H1 AOS
vladimir.bossilkov@LIGO.ORG - posted 17:23, Wednesday 22 January 2020 - last comment - 09:49, Thursday 23 January 2020(54662)
Recreating systematic errors in Calibration by deliberately altering the model with Kappas at the time of measurement

I am experimenting in trying to manually multiply the all the Kappas and set the cavity pole as at the time of measurement on 20200113, as is being discussed in the last two comments in this aLOG thread.

 

I used ndscope to acquire Kappas for: UIM, PUM, TST, Optical Gain, and Cavity Pole (not really Kappa, just altered the actual value to the one at time of measurement).

Attached here are two plots:

I am beginning to suspect that the estimate of the Kappa_PUM might be grossly exaggerated? Perhaps the frequency where it is being sensed is not a great choice?

Images attached to this report
Comments related to this report
vladimir.bossilkov@LIGO.ORG - 09:49, Thursday 23 January 2020 (54671)

Jeff is currently working hard to get a better figure for the actuation force coefficients. However if what I am suggesting above is true: if he improves the actuation force coefficient for the PUM stage, then the Kappa_PUM would change by the magnitude of the change he makes. At the end of the line, after applying Kappas there would actually be no change at all, and this problem would perpetuate.