Reports until 14:11, Tuesday 28 January 2020
H1 CDS
david.barker@LIGO.ORG - posted 14:11, Tuesday 28 January 2020 - last comment - 16:04, Tuesday 28 January 2020(54779)
CDS Maintenance Summary, Tuesday 28th January 2020

WP8520 Adding two fast channels to LSC model

Sheila, Dave:

Sheila made a new h1lsc model which adds two new 16kHz channels to the DAQ:

h1lsc was restarted followed by a DAQ restart

WP8521 Add SUS SW2R channels to DAQ for ramping trends

Dave:

I created a new H1EPICS_SUSSW2R.ini file and added it to the DAQ.

DAQ Restart, Green up EDC

Dave:

DAQ was restarted for three changes:

  1. h1lsc fast channels added
  2. H1EPICS_SUSWS2R slow channels added
  3. H1EPICS_PSLOPC0 slow channels removed while h1pslctrl0 IOC is non-operational

Control Room FOMs

Dave:

I patched the OS on all the control room FoM machines and rebooted them.

Enabled tagging on EX SEI and SUSAUX network ports

Dave:

I enabled ADMINLAN tagging to the H1FELAN ports of h1seiex and h1susauxex to try to get remote IPMI working again for these machines. This was a partial success, I can now ping the IPMI addresses for these machines but ipmi-tool does not connect to them. We may need to power cycle the computers or cold-cycle the IPMI cards to complete this.

Comments related to this report
david.barker@LIGO.ORG - 16:04, Tuesday 28 January 2020 (54782)

New (Unify) and old (Cisco) WAP status are being shown side-by-side on the CDS Overview MEDM while we are in this transition period. Unify are shown a squares, Cisco as circles.

Images attached to this comment