jenne.driggers@LIGO.ORG - posted 19:15, Wednesday 29 January 2020 (54799)
SRY locking thershold lowered to 0.3

[Camilla, Jenne, Sheila]

Lowered SRCL thresh.

I had seen this while sitting with Jim during an initial alignment earlier this week, and Sheila noted that she'd seen it recently too, but we're not getting the LSC POP DC power above the old ON threshold value of 0.4, so SRY never triggers and sends control signal to the SRM, so we keep getting stuck in the Acquire_SRY state of initial alignment.  I have changed the threshold in lscparams for SRXY locking to turn on at 0.3 and off at 0.2 (from the former values of on at 0.4 and off at 0.3).