Displaying report 1-1 of 1.
Reports until 09:24, Thursday 13 February 2020
H1 General
thomas.shaffer@LIGO.ORG - posted 09:24, Thursday 13 February 2020 - last comment - 13:04, Friday 20 March 2020(55081)
Lock Loss 1723 UTC

No obvious cause.

Comments related to this report
thomas.shaffer@LIGO.ORG - 09:42, Thursday 13 February 2020 (55082)Lockloss, PSL

The FSS started to oscillate about 150seconds 30seconds before we lost lock. During relocking we keep losing lock at various places and there is slight motion seen in the Ref Cav Trans spot.

Images attached to this comment
thomas.shaffer@LIGO.ORG - 09:58, Thursday 13 February 2020 (55083)Lockloss, PSL

The lock loss at 0714 UTC looks similar as well. Something is seen in that channel 45 seconds before.

Images attached to this comment
camilla.compton@LIGO.ORG - 16:57, Wednesday 19 February 2020 (55188)Lockloss
I have looked into what are normal ranges for this channel are when we are locked (over the past month and a snapshot in July, which looked similar).
It seems that we survive periods with the FSS piezo channel H1:PSL-FSS_FAST_MON_OUT_DQ  sustained near 3. And survive fast spikes up 9. Seen in the attached images (sustained vs peaks).
The value of this channel on both cases the IFO lost lock that TJ found, were sustained near 6. so much higher than we normally see.
Images attached to this comment
camilla.compton@LIGO.ORG - 13:04, Friday 20 March 2020 (55697)PSL
In the week spoken about above, we had 5 locklosses that looked like this with the FSS_NPRO_TEMP and FSS_FAST (PZT) signals oscillating shortly before lockloss:
2020-02-15 10:41:46 UTC,  2020-02-15 04:11:04 UTC, 2020-02-14 14:13:51 UTC,  2020-02-13 17:23:22 UTC, 2020-02-13 07:14:20 UTC
We then had no more locklosses that looked like this until this week (1 month on), where we have had 3:
2020-03-19 00:02:17 UTC, 2020-03-17 14:25:22 UTC (image attached), 2020-03-17 12:12:47 UTC. 
Microseism was high in the first set of locklosses, but this week all the environmental conditions have been calm.
We should continue to investigate what signal is becoming unstable and being fed back into the PSL signals or whether the PSL is causing this.
Images attached to this comment
Displaying report 1-1 of 1.