TITLE: 02/22 Day Shift: 16:00-00:00 UTC (08:00-16:00 PST), all times posted in UTC STATE of H1: Observing at 120Mpc INCOMING OPERATOR: Ed SHIFT SUMMARY: I was told by Cheryl that we are keeping INJ_TRANS set to INJECT_KILL for the indefinite future due to the possibility of a supernova. There was one lock loss from observing. I ran across a user code error in CHECK_MICH_FRINGES. I think I may have fixed it, but CHECK_MICH_FRINGES did not leave the alignment in a good state for ACQUIRE_PRMI. We lost lock at ACQUIRE_PRMI twice before I ran an initial alignment. The initial alignment seemed to fix the issue. Currently riding through a series of earthquakes in Chile with SEI_CONF in EARTH_QUAKE mode. LOG: 20:01 UTC Lock loss 20:29 UTC Lock loss from MOVE_SPOTS Fixed user code error in CHECK_MICH_FRINGES (alog 55231) 20:58 UTC Lock loss from ACQUIRE_PRMI 21:16 UTC Lock loss from ACQUIRE_PRMI Starting initial alignment 21:43 UTC Initial alignment done 22:01 UTC Lock loss from ENGAGE_ASC_FOR_FULL_IFO 22:43 UTC Observing. No SDF differences. 23:22 UTC Set SEI_CONF to EARTH_QUAKE for magnitude 6.0 earthquake in Chile