Reports until 11:45, Monday 24 February 2020
H1 CAL (CAL, DetChar)
sudarshan.karki@LIGO.ORG - posted 11:45, Monday 24 February 2020 - last comment - 14:01, Tuesday 25 February 2020(55260)
PCALX Laser Issues

Evan and Yanyan noted that PCALX was malfunctioning since Wednesday (LHO alog 43959 and 55257). Jeff has added an alert that will notify operators if this happens in the future (LHO alog 55257). 

Since, simply switching on and off the OFS control loop, didnot bring PCAL back, I looked at the PD singals and saw all PD signals were reading zero values. I tried switching on and off the laser remotely but still no signals on any of the photodiodes. Further investigations (on AOM status and/or the laser itself) will require  going down to the endstation. Stay tuned!

 

Comments related to this report
sudarshan.karki@LIGO.ORG - 17:54, Monday 24 February 2020 (55275)CAL, DetChar

The Pcal laser could have been potentially shut off because of the interlock system that has been down at X-end since Tuesday maintenance (last week). Someone will have a look at it tomorrow to see if that was the cause.

jeffrey.kissel@LIGO.ORG - 10:55, Tuesday 25 February 2020 (55286)CDS, DetChar, INJ, ISC, OpsInfo
For more details of what we believe happened, see comment LHO aLOG 55285.

Stay tuned for results of today's investigation.
jeffrey.kissel@LIGO.ORG - 14:01, Tuesday 25 February 2020 (55294)CAL, CDS, DetChar, INJ, OpsInfo
PCALX Laser functionality restored after interlock was repaired today. See LHO aLOG 55291.