Reports until 08:19, Wednesday 26 February 2020
H1 OpsInfo (OpsInfo)
edmond.merilh@LIGO.ORG - posted 08:19, Wednesday 26 February 2020 - last comment - 08:38, Wednesday 26 February 2020(55311)
INJ_TRANS not Changing to INJECT_KILL for GRB?

15:30UTC I arrived on site.

15:35 I heard Verbal Alarms shouting about GRB Short - E365598 Tig Dur .016 and acknowledged

Anyway, those are my observations from this morning and food for thought regarding our current trial.

 

Images attached to this report
Comments related to this report
thomas.shaffer@LIGO.ORG - 08:38, Wednesday 26 February 2020 (55312)

Injections could come unannounced, so we need to be prepared for that.

INJ_TRANS will not go to INJECT_KILL when an external alert comes in, but it will automatically go to EXTTRIG_ALERT_ACTIVE as is shown in your screenshot of the log. This will do basically the same thing as INJ_KILL and will stay in that state for an hour. We manually take this node to INJ_KILL on GRBs for redundancy, making sure that there is no chance of an injection happening at that time.

The stand down time also is for site operations. We need to limit any activity on site, and make sure that maintenance and commissioning activities are delayed until after the stand down.

(I should change Verbal to handle the trial better).