(Gray, Kissel, Thomas)
Patrick started this measurement at the end of his shift and I completed the measurement. Below is a summary:
- Start: 23:10:53 (1266966671 gps time): H1 reaches INJECT SQUEEZING (but has oscillations) (from Patrick's alog)
- [at time above?] Patrick starts Cal Line Comb Test (I believe before H1 was in NLN due to the squeezer oscillations).
- 23:59:52 (1266969610 gps time): H1 reaches NOMINAL LOW NOISE
- Stop: 1:17:00 (1266887838gps) aborted (i.e. "stopped") the measurement. (per alog 55297 & with Jeff on phone)
- Then took screen shot of spring frequency squared (attached)
- Reverted both CALEX & CALEY sdfs.
- Then set PCALx line to hits most recent line frequency value.
- Additionally confirmed everything's good by looking at:
- PCALE[x/y] medms: OFS PD "strip tool" nice & flat around -5, Error Monitor signal was around 0.
- DARM spectra on wall: Pcal[x&y] lines at all the desired frequencies.