[Sheila, Jenne]
Although the squeezer seems to not have gone into oscillation since last Thursday when we lowered the green power (alog 55339), we are looking to see if we can better understand the cause and ameliorate it (spoiler alert: so far we don't know, although we have further tests to try today).
Here's what we've done so far:
Jenne cabled the LO PZT driver channel (used only for locking with the diagnostic homodyne, not during runing) to the OPO PZT2. The attached screenshot shows that with the OPO locking using PZT1 (blue line, around 50V), we can move the offset on PZT2 (LO channel, blue line around 0-20V) to reduce the voltage on PZT1. You can see that the reflected power increases when there is a higher voltage on PZT2, which is on one of the curved mirrors, because the cavity gets more misaligned from driving PZT2 than PZT1.
We tried but weren't able to recreate the oscillation a second time, so we don't know if moving around the offset on the second PZT will have any impact on the instability. We can try to move this offset if the instability does reappear, although it might not help if the problem is related to the misalignment of the OPO that the high PZT1 voltage causes.
Keita was wondering if part of the difference (that prevented us from recreating the oscillation) was the fact that both PZTs were connected to cables, neither of them was connected to a terminator. So, I reverted PZT2 back to being a terminator (and re-plugged in the LO cable to the LO PZT output of the driver). Now everything cable-wise is back to how it was this morning. However, I still wasn't able to cause oscillations, even if I brought the green power up above 1.8.
So, for now I'll leave all of the cables as they are, which is fully reverted from the tests today. If we see the oscillations come back, then we'll reconsider again borrowing the LO driver channel for PZT2.
When I left things at a green power of 1.3, I had also reverted the squeezer angle, but had forgotten to revert the OPO crystal temperature, H1:SQZ-OPO_TEC_SETTEMP. I have now reverted it to its value from before last Thursday's green power change (33.014 C). I did not measure the nonlinear gain. For this, we were out of Observing for ~1 minute.