FAMIS 11055
Laser Status:
Front End Power is 31.93W (should be around 30 W)
70W Output Power is 70.3W
Front End Watch is GREEN
70W Watch is GREEN
PMC:
It has been locked 4 days, 1 hr 26 minutes (should be days/weeks)
Reflected power = 11.86Watts
Transmitted power = 51.77Watts
PowerSum = 63.63Watts.
FSS:
It has been locked for 1 days 19 hr and 23 min (should be days/weeks)
TPD[V] = 5.586V (min 0.9V)
ISS:
The diffracted power is around 2.5%
Last saturation event was 1 days 19 hours and 22 minutes ago (should be days/weeks)