H1 CAL
jeffrey.kissel@LIGO.ORG - posted 13:39, Monday 16 March 2020 (55624)
Calibration Measurements Today: Full Suite, after OMC Whitening Chassis Filters Reduced from 2 stages of Whitening + 1 Low Pass to Just 1 Whitening
J. Kissel

I've gathered the full, regular calibration measurement suite today -- importantly -- *after* the OMC whitening chassis' filter configuration change (see LHO aLOG 55620).

These measurements may end up being the reference measurements for a future calibration model parameter set.

Data is committed here 
    /ligo/svncommon/CalSVN/aligocalibration/trunk/Runs/O3/H1/Measurements/FullIFOSensingTFs/
        2020-03-16_H1_DARM_OLGTF_LF_SS_5to1100Hz_15min.xml
        2020-03-16_H1_PCALY2DARMTF_LF_SS_5t1100Hz_10min.xml

        2020-03-16_H1_PCALY2DARMTF_BB_3min.xml    << Contains before (reference) and after (live) OMC whitening filter change. 
              "Before change" measurement started at 2020-03-16 18:30:31 UTC, 
              "after changes" started at 2020-03-16 18:35:14 UTC.

    /ligo/svncommon/CalSVN/aligocalibration/trunk/Runs/O3/H1/Measurements/FullIFOActuationTFs/
        2020-03-16_H1SUSETMX_L1_iEXC2DARM_8min.xml
        2020-03-16_H1SUSETMX_L1_PCAL2DARM_5min.xml
        2020-03-16_H1SUSETMX_L2_iEXC2DARM_12min.xml
        2020-03-16_H1SUSETMX_L2_PCAL2DARM_6min.xml
        2020-03-16_H1SUSETMX_L3_iEXC2DARM_12min.xml
        2020-03-16_H1SUSETMX_L3_PCAL2DARM_6min.xml

Analysis to come.