J. Kissel [remotely]
I've gathered the standard, full, suite of calibration measurements today. The full suite will be useful for building up the new reference model that we'll likely need after the OMC whitening chassis configuration change last week. Other than the new OMC whitening chassis configuration, the IFO was otherwise operating normally for O3B. ~38W input to the IMC, 50 ct SRCL offset, usual configurations of the actuator.
Raw data files have been committed to the following directories:
/ligo/svncommon/CalSVN/aligocalibration/trunk/Runs/O3/H1/Measurements/FullIFOSensingTFs/
2020-03-23_H1_DARM_OLGTF_LF_SS_5to1100Hz_15min.xml
2020-03-23_H1_PCALY2DARMTF_LF_SS_5t1100Hz_10min.xml
2020-03-23_H1_PCALY2DARMTF_BB_3min.xml << injection started at 2020-03-23 18:01:20 UTC
/ligo/svncommon/CalSVN/aligocalibration/trunk/Runs/O3/H1/Measurements/FullIFOActuationTFs/
2020-03-23_H1SUSETMX_L1_PCAL2DARM_5min.xml
2020-03-23_H1SUSETMX_L1_iEXC2DARM_8min.xml
2020-03-23_H1SUSETMX_L2_PCAL2DARM_6min.xml
2020-03-23_H1SUSETMX_L2_iEXC2DARM_12min.xml
2020-03-23_H1SUSETMX_L3_PCAL2DARM_6min.xml
2020-03-23_H1SUSETMX_L3_iEXC2DARM_12min.xml
Data analysis to come.