h1pemmx front end crash

The h1pemmx system crashed at 13:27 UTC (06:27 PDT) this morning, Thursday 2nd April 2020. It does not respond to pings and will need to be power cycled to restore it. Talking with Richard, because this system reads out the Vault PEM signals, we will want to reboot this at the earliest opportunity, which could be next Tuesday when team-vacuum is onsite.

On the CDS Overview MEDM: h1ioppemmx and h1pemmx models are white, h1edc is not connecting to 17 channels for these models (9 for iop, 8 for pemmx) and the CDS Hardware status is RED with 'ioppemmx ADC error'.

If the downtime will be many days, I'll change these sytems to bypass pemmx to 'green them up'.

The sister system pemmy has been running for 561 days (since 19th September 2018, last RCG upgrade), presumably h1pemmx had the same uptime.

Fast channels missing while this system is down (tagging PEM and DetChar)

[H1:PEM-MX_ACC_BEAMTUBE_CRYO_Y_DQ]
[H1:PEM-MX_SEIS_VEA_FLOOR_QUAD_SUM_DQ]
[H1:PEM-MX_SEIS_VEA_FLOOR_X_DQ]
[H1:PEM-MX_SEIS_VEA_FLOOR_Y_DQ]
[H1:PEM-MX_SEIS_VEA_FLOOR_Z_DQ]
[H1:PEM-VAULT_MAG_1030X195Y_COIL_QUAD_SUM_DQ]
[H1:PEM-VAULT_MAG_1030X195Y_COIL_X_DQ]
[H1:PEM-VAULT_MAG_1030X195Y_COIL_Y_DQ]
[H1:PEM-VAULT_MAG_1030X195Y_COIL_Z_DQ]
[H1:PEM-VAULT_SEIS_1030X195Y_STS2_QUAD_SUM_DQ]
[H1:PEM-VAULT_SEIS_1030X195Y_STS2_X_DQ]
[H1:PEM-VAULT_SEIS_1030X195Y_STS2_Y_DQ]
[H1:PEM-VAULT_SEIS_1030X195Y_STS2_Z_DQ]
 

Images attached to this report