SUS health checks - transfer function of double sus: OMC, TMSX, TMSY

Cheryl, Rahul

Attached below are the transfer function measurements for 3 double suspensions namely, OMC, TMSX and TMSY. They are healthy and clear of any rubbing. We also started with single suspensions and were allmost done with RM1 and RM2, however we found some issues in the reference plot. We will post RM1 and RM2 results tomorrow morning.

The templates are stored at the following locations,

/ligo/svncommon/SusSVN/sus/trunk/OMCS/H1/OMC/SAGM1/Data/

2020-04-16_1500_H1SUSOMC_M1_WhiteNoise_L_0p02to50Hz.xml
2020-04-16_1500_H1SUSOMC_M1_WhiteNoise_P_0p02to50Hz.xml
2020-04-16_1500_H1SUSOMC_M1_WhiteNoise_R_0p02to50Hz.xml
2020-04-16_1500_H1SUSOMC_M1_WhiteNoise_T_0p02to50Hz.xml
2020-04-16_1500_H1SUSOMC_M1_WhiteNoise_V_0p02to50Hz.xml
2020-04-16_1500_H1SUSOMC_M1_WhiteNoise_Y_0p02to50Hz.xml

/ligo/svncommon/SusSVN/sus/trunk/TMTS/H1/TMSX/SAGM1/Data

2020-04-16_1500_H1SUSTMSX_M1_WhiteNoise_L_0p02to50Hz.xml
2020-04-16_1500_H1SUSTMSX_M1_WhiteNoise_P_0p02to50Hz.xml
2020-04-16_1500_H1SUSTMSX_M1_WhiteNoise_R_0p02to50Hz.xml
2020-04-16_1500_H1SUSTMSX_M1_WhiteNoise_T_0p02to50Hz.xml
2020-04-16_1500_H1SUSTMSX_M1_WhiteNoise_V_0p02to50Hz.xml
2020-04-16_1500_H1SUSTMSX_M1_WhiteNoise_Y_0p02to50Hz.xml

/ligo/svncommon/SusSVN/sus/trunk/TMTS/H1/TMSY/SAGM1/Data

2020-04-16_1800_H1SUSTMSY_M1_WhiteNoise_L_0p01to50Hz.xml
2020-04-16_1800_H1SUSTMSY_M1_WhiteNoise_P_0p01to50Hz.xml
2020-04-16_1800_H1SUSTMSY_M1_WhiteNoise_R_0p01to50Hz.xml
2020-04-16_1800_H1SUSTMSY_M1_WhiteNoise_T_0p01to50Hz.xml
2020-04-16_1800_H1SUSTMSY_M1_WhiteNoise_V_0p01to50Hz.xml
2020-04-16_1800_H1SUSTMSY_M1_WhiteNoise_Y_0p01to50Hz.xml

Images attached to this report