"Low level - Warning" in Diode Room Ante-chamber (room 172B)

FYI - I noticed an audible "beep" coming from the chiller unit in the ante-chamber room (room 172B) adjacent to the OSB receiving area. 

Comments related to this report
kyle.ryan@LIGO.ORG - 09:27, Tuesday 21 April 2020 (55938)

At Jeff B.'s direction I added 200 ml of water which silenced the alarm.  He will be in tomorrow, as normal, and will check. 

jeffrey.bartlett@LIGO.ORG - 09:32, Tuesday 21 April 2020 (55939)
   Had Kyle (who was on site for N2 deliveries) add 200ml water to crystal chiller. This will keep it running until I can get in and give both the chillers a full check out. 

   Looks like the Crystal chiller will run for about 3 weeks on a full fill before getting close to a low water shut down.