Displaying report 1-1 of 1.
Reports until 16:29, Friday 15 March 2013
H1 SUS
jeffrey.kissel@LIGO.ORG - posted 16:29, Friday 15 March 2013 (5823)
H1 SUS ITMY
Was poking around H1 SUS ITMY to try and assess SUS sensor noise, and found that ITMY needed some infrastructure TLC similar the HAM SUS. 
- CART2EUL and EUL2CART matrices were still using H2 values, so I've updated them
- OSEMINF to_um calibration in FM5 was old, stale, and off. It's now correct, fresh, and ON.
- Stuck from_um anti-calibration filters is the DAMP filter banks such that we can use the same loops.
 
I've since captured a new safe.snap, and committed both the new filter file and h1susitmy_safe.snap to the userapps repo.