Displaying report 1-1 of 1.
Reports until 22:00, Monday 22 March 2021
H1 SQZ
sheila.dwyer@LIGO.ORG - posted 22:00, Monday 22 March 2021 - last comment - 11:00, Tuesday 23 March 2021(58323)
OPO reflection path mode

Camilla and Corey have replaced the green mirror in reflection that was scratched (this was an optic reused from O3).  This afternoon with the new optic in place I returned to finshing the G path, in reflection off the OP towards SQZT7.

The first attachment shows the measurement with no lenses between the OPO and the beam profiler. The vertical waist is 175um radius 278mm before G:M2 (which should be about 65mm from OPO M1).  The horizontal waist is 204 um 201 mm before G:M2 (meaning 142 mm from OPO M1). 

After placing the +150mm ROC lens about 50mm from G:M2, I remeasured the beam shown in second attachment. The resutling vertical waist of 323um is 364mm downstream of G:M2; horizontal waist of 263um is 248mm downstream of G:M2.

I didn't take care to block the OPO while making these measurements, which would have been a good idea to prevent passing through a resonance from disturbing the measurement. These measurents aren't in agreement with T1900649-v8 figure 22.  

Images attached to this report
Non-image files attached to this report
Comments related to this report
lee.mcculler@LIGO.ORG - 11:00, Tuesday 23 March 2021 (58333)SQZ

Ok, I looked back at my mode matching code. In that case it was calculating along an incorrect path for an obvious reason that I missed given all the many paths setup in that document. I fixed the issue and attached is a new Fig 22 Fig22_improved.png. It agrees much better with your measurements and indicates that you can keep G:L3 where it is. I also checked how the matching with the SK path changes. It also improves quite a bit, since the new PUMP-REFL beam is better collimated. The matching with the FCGS -> SK path beam is 91\%.

 

With this fired up, I'm looking into my OPO model to also look at the A: path measurements. I'm getting better matching to the A:L1-A:l3 measurements after some adjustments. I'll post more later, since I'll probably have to adjust the layout a bit to ensure good matching.

Images attached to this comment
Displaying report 1-1 of 1.