Displaying report 1-1 of 1.
Reports until 12:50, Friday 22 March 2013
H1 SUS
mark.barton@LIGO.ORG - posted 12:50, Friday 22 March 2013 (5864)
BS safe.snap link
Betsy noticed that the BS screen was in a bad state, with gains and offsets and matrix values not filled in, despite the fact that I'd done setup on it back on 2/18 (alog 5440). It turns out it was because although I'd created 

/opt/rtcds/userapps/release/sus/h1/burtfiles/h1susbs_safe.snap

there was no symbolic link

/opt/rtcds/lho/h1/target/h1susbs/h1susbsepics/burt/safe.snap 

pointing to it. So after the various restarts between now and then it was not getting correctly restored. I created the link and checked that the correct settings were loaded when I restarted the model.
Displaying report 1-1 of 1.