H1 ISC
christopher.wipf@LIGO.ORG - posted 00:32, Monday 25 March 2013 (5873)
End Y slow controls signals jumbled
(Sheila, Max, Chris)

On Friday we uncovered a snafu with the slow controls cabling at the Y end.  The connector labels engraved on the rear panel of the "End 2" EtherCAT chassis (Exhibit A) do not agree with the chassis wiring plan (Exhibit B).  Upon opening the chassis it appeared that the internal wiring to the Beckhoff modules was done according to (B), whereas the external cable runs had been connected relying on (A).  Here's a table showing the correspondences:
PortWiring planRear panel description
6AuxiliaryDC Photodiodes
7PZT Mirrors/PDs/Laser TempAuxiliary
8TCSPZT Mirrors/PDs/Laser Temp
9RF AmplifiersTCS
10DC PhotodiodesRF Amplifiers
11LSC DemodulatorsLSC Demodulators
12Laser DiagnosticsLaser Diagnostics
This problem came to our attention after we noticed that the fiber PLL slow path (laser temperature) control output was broken. Actually it was just routed onto the wrong cable. Attached images show the interior and rear panel of the End 2 chassis at the Y end.