WP9880
I removed the O3 ZM2 tip tilt from HAM5 and moved the SRM heater optics to be used in the upcoming temporary laser setup for OFI alignment.
Summary of activities:
- Dust counts were 0's or low 10's before and after opening soft covers, using grey handheld
- Collected 4" and 1" CC samples with Nidhi's help
- Removed cables, BOSEMs, and optic from ZM2
- Corey and I bagged the TT, BOSEMs, TT spacer, beam dump spacer, dog clamps
- I left 3 beam dumps and the SRM heater mirrors on the South side of the table to be, potentially, used for the upcoming temp. laser setup.
just for info - the optic used for ZM2 (HTTS, HAM5) optic is E1000425 - currently stored at LHO.