Follow up to alog 60633. WP 10035. Procedure T2100117 and cabling doc D1101814.
TJ, Austin, Camilla. Thursday 11th November PM.
Summary: All picomotors in and tested. TCS in chamber work is done. Need to replace the in-air 25pin cable before functional.
Other: Jim ran the CB to flange cable but it has no captured screws in on the CB side. I added 4-40's to attach it both cables to each side of the cable bracket so it is secure. Normally the screws to attach to other cable housing would have been added to the cable in production so not easy to add now. Photo here, the top side 40-4's look part way in but there are threaded 4-40's so holding their cable to the bracket fine.
Controller channel | x | y | |
#1-4 | in-air HWS table controls for periscope mirrors. | - | - |
#5 | HWSY M1 | Yaw: +x = SR2 refelcted beam moves in +x | Pitch: + y = up |
#6 | HWSY M2 | Yaw: +x = SR2 refelcted beam moves in -x | Pitch |
#7 | HWSX M2 | Yaw | Pitch |
#8 | HWSX M1 | Yaw | Pitch |
All worked as expected. Where only Yaw or Pitch is written we tested the pico was working be feeling and hearing it move but didn't sett a laser up to see the direction of movement.
I cannot get the advanced screens or the singe motor screens to open via medm from the PICO_H screen. I wonder if this is a similar problem to the rcg one Jonathon/Dave fixed last week (alog 60552). Tagging cds.
Daniel fixed the picomotor medm screens and HWS HAM4 picomotor names. WP#10055
We also discovered that the tcs system has lost all of its filters. This happened when we added a top_names block for TCS, in order to support a separate AWC top name. Even so no channel name changed, the names of the filters in the filter file did. For instance, ITMX_RH_INVERSE_FILTER became TCS_ITMX_RH_INVERSE_FILTER, etc. Since RCG only updates the module name list and not the actual filters, the old filters were removed next time the file was saved with foton. The old filters are now restored with the updated names.
Fil replaced the in-air cable and all picos have now been tested with correct cabling: alog 60674